gongjigongyi.com

与米乐m6网页版登录入口代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

What is the time required to switch between Pulse IV (4225PMU) and CV (4210ACVU) measurements using the 4225-RPM for the 4200A-SCS?

问题:

What is the time required to switch between Pulse IV (4225PMU) and CV (4210ACVU) measurements using the 4225-RPM for the 4200A-SCS?

答案:

The RPM eliminates the need to re-cable and increases switching speed between Pulsed IV and CV measurements.  However, we do not specify the switching time for the RPMs in the data sheets.

A simple test using a MDO3102 yeided the below approximate switch times:
CVU to PMU measurements through RPM = 460ms.
PMU to CVU measurements through RPM = 1700ms.


此常见问题适用于:

产品系列: Keithley 4200A-SCS 参数分析仪

产品:

常见问题 ID 469646

查看所有常见问题 »