gongjigongyi.com

与米乐m6网页版登录入口代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

The data sheet for my oscilloscope says that the vertical input sensitivity is 1 mV/div to 10 V/div in 1 MΩ mode. But I turn the Vertical Scale control fully clockwise and I can only get down to 10 mV/div. What am I doing wrong?

问题:

The data sheet for my oscilloscope says that the vertical input sensitivity is 1 mV/div to 10 V/div in 1 MΩ mode. But I turn the Vertical Scale control fully clockwise and I can only get down to 10 mV/div. What am I doing wrong?

答案:

You probably aren't doing anything wrong. But you do have a 10X probe attached to that channel, right?

The 1 mV/div specs relate to the oscilloscope inputs. If the signal’s source impedance is relatively low (<< 1 MOhm), the signal’s frequency is relatively low (less than a few MHz), and the voltage is within the input range of the oscilloscope, you can connect the signal directly to the oscilloscope with a cable and the vertical scale does go down to 1 mV/div in 1 MOhm mode.

However, when you attach an attenuating probe, the mV/div scale factor always increases by that same factor. This is a standard feature of oscilloscopes that automatically sense the probe attenuation.

Learn the XYZ's of Oscilloscopes »


常见问题 ID 58571

查看所有常见问题 »